Enabling better global research outcomes in soil, plant & environmental monitoring.

Lab Setup 0700LAB023V2

The Model 0700LAB023V2 is a multiple porous plate extractor system for laboratory use. This is the most popular lab system, providing capabilities for a variety of soils and conditions. The 1500 and 1600 extractors together provide ample flexibility and capacity for low or high water content extraction routines. The combination provides for 2 extractor units for optimum plant growth studies (0-5 Bar Water Retention Studies) while having 15 Bar capabilities as well.

1500F2 Specifications

Pressure Range 15 Bars
Pressure Vessel Inner Diameter 30.48cm (12″)
Depth 10.16cm (4″)
Number of Samples ~36 x 5.715cm diameter samples (2.25″)
Dimensions 48.26 x 48.26 x 40.64cm (19 x 19 x 16″)
Weight 41 kg (90 lbs)

 

1600F1 Specifications

Pressure Range 0-5 Bar
Pressure Vessel Inner Diameter 30.48cm
Depth 22.86cm (9″)
Number of Samples ~48 x 5.715cm diameter samples
Weight 18.6 kg (41 lbs)

The Lab Setup 0700LAB023V2 is comprised of:

  • 0700CG23F1 Manifold, a combination manifold which combines the 0700G2 station and the 0700G3F1 station. The 0700G2 station is used to operate both the 1500F2, 15 Bar Ceramic Plate Extractor and the 1600F1, 5 Bar Pressure Plate Extractor. The 0700G3F1 station provides very precise, low-pressure regulation in the pressure range from 1 to 75 psi.
  • 1500F2, 15 Bar Pressure Plate Extractor, used to analyse the water-holding characteristics of soil samples throughout the pressure range of interest in most agricultural applications. The pressure vessel is 4″ (10 cm) deep and has an inside diameter of 12″ (30 cm). Up to 4 ceramic plates can be accommodated at one time, allowing approximately 36 2-1/4″ samples to be analyzed simultaneously. The Pressure Plate Extractor consists of a pressure vessel and lid, clamping bolts, O-ring seals, and outflow tube assemblies.
  • 1600F1, 5 Bar Pressure Plate Extractor, allows analysis of the water-holding characteristics of soil samples in the 0 to 5 bar pressure range. The pressure vessel is 9″ (22 cm) deep and has an inside diameter of 12″ (30 cm). Up to 4 ceramic plates can be accommodated at one time, allowing approximately 48 each 2-1/4″ (5.7 cm) diameter samples to be analysed simultaneously. The Pressure Plate Extractor consists of a pressure vessel and lid, clamping bolts, O-ring seals, and outflow tube assemblies.
  • 4x 0675B15M1, 15 Bar Extractor Pressure Plate Cell. Developed from a proprietary mixture of ball clays fired to a ceramic body. Pinkish-tan in colour, moderately hard and will withstand pressure differentials of 15 Bar (217.5 psi). Typical bubbling pressure: 220 psi. Effective pore size: 0.16 micron. Hydraulic conductivity: .00000000259 cm/sec. Approximate porosity: 32% by vol. Flow through
  • 4x 0675B05M1, 5 Bar Extractor Pressure Plate Cell. Recommended for specialised applications where the pressure differentials are under 5 Bar (72.5 psi). Typical bubbling pressure: 80 psi. Effective pore size: 0.5 micron. Hydraulic conductivity: .000000121 cm/sec. Approximate porosity: 31% by vol. Flow through
  • 4x 0675B03M1, 3 Bar Extractor Pressure Plate Cell. A complex mixture of ball clays into a moderately fired ceramic body. Good porosity and good hydrologic flow capability. The material is moderately hard and tannish-white in colour, not recommended for fluid sampling. Recommended for specialised applications where the pressure differentials are under 3 Bars (43.5 psi). Typical bubbling pressure: 46 – 70 psi. Effective pore size: 0.7 microns. Hydraulic conductivity: .00000025 cm/sec. Approximate porosity: 34% vol. Flow through
  • 4x 0675B01M3, 1 Bar Extractor Pressure Plate Cell. Developed from a high fired Alumina body. Extremely porous, inert to most all solutions, possesses hard exterior and interior surfaces, and pure white in color. Recommended for standard pressure differentials under 1 Bar (14.5 psi). Typical bubbling pressure: 28 – 32 psi. Effective pore size: 2.5 microns. Hydraulic conductivity: .0000086 cm/sec. Approximate porosity: 45% by vol. Flow through
  • 1093F1 Soil Sample Retaining Rings. Eight sets of 12, used to hold soil samples on a pressure plate cell during an extraction run.
  • 0775L60 Connecting Hose. 60″ (152.4cm) long, for 1500 extractor. 1/2″ O.D., 1/4″ I.D. pressure tubing, with swivel fittings to connect to the manifold and extractor.
  • 0776L60 Connecting Hose. 60″ (152.4cm) long, for 1600F1 extractor. 1/2″ O.D., 1/4″ I.D. pressure tubing, with a swivel fitting to connect to the manifold, and a hose clamp to connect to the extractor.
  • 0779G1 Connecting Hose. For connecting compressor to manifold. 1/4″ male pipe to standard hose valve, and a 1/4″ NPTM x 1/4″ NPTF brass street elbow.
  • 1080G1 PM Hinge. For use with the 1500F2 extractor. The hinge attaches to the back of the extractor, for ease of opening.
  • 1055K1 Right Angle Outflow Adapter Kit. Set of 4. 3/16″ O.D. Neoprene tubing, and 3/32″ O.D. Nylon tubing. Useful for modifying the outflow path from extractors.
  • 1057K1 Plug Bolt with Seal. Set of 5 stainless steel plug bolts, 1/4″ diameter threaded male fitting.
  • 0505V2, 240V Compressor.

 Optional:

  • 0675B0.5M2, 0.5 Bar high flow pressure plate. High fired, Alumina body. An excellent material, extremely porous, and inert to most all solutions. Hard exterior and interior surfaces and pure white in color. Recommended for low air pressure differentials not exceeding 0.5 Bar (7.25 psi). Typical bubbling pressure: 12-16 psi. Effective pore size: 6.0 microns. Hydraulic conductivity: .0000311 cm/sec. Approximate porosity: 50% by vol. Typical flow through for a 1/4-inch plate: 180 ml/hr/cm2 at 14.7 psi. Ideal for liquid or gas sampling as the Alumina material has almost no ionic exchange sites or leachable mineralogy. Truly a superior ceramic for both industrial and scientific work where high volume transfer or testing at low pressure differentials is needed.
  • 0675B01M1, 1 Bar pressure plate.